Producción científica2015· International journal of odontostomatology

Accuracy of Resonance Frequency Analysis by Third Generation Osstell®

Q4 · Health ProfessionsImpact Factor · 0.197

AutoresAritza Brizuela‐Velasco, Felipe J. Fernández-González, Nerea Martín-Blanco, David Chávarri‐Prado, Yelko Chento-Valiente

Abstract

The Resonance Frequency Analysis by Osstell is used in the assessment of primary and secondary implant stability. The aim of this study was to determine its accuracy as a method of testing, checking its repeatability and reproducibility. A pilot study was performed whereby implants were placed in a cow rib to compare several ISQ measurements with the same SmartPeg transducer (repeatability), or with different SmartPegs (reproducibility). Statistical analysis of the data showed a high correlation of the ISQ values obtained in both assays, which suggests that Osstell analysis can provide high repeatability and reproducibility.